Frequently Asked Questions
What is the recommended liquid or solvent with which to clean the sample cell?
If I'm not getting good readings from my analyzer, what should I look at first?
How can we change the range on a 7600 analyzer? Only a small change is required from 0-50 ppm CO to 0-60 ppm.
For additional questions, please contact Technical Support.
Q: What is the recommended liquid or solvent with which to clean the sample cell?
A: It is recommended to use ethanol (ethyl alcohol) with purity of at least 99.5%, followed by flowing dry nitrogen once the cell is re-installed.
Back to Top
Q: If I'm not getting good readings from my analyzer, what should I look at first?
A: Make sure the purge gas is connected and running into the analyzer.
Q: How can we change the range on a 7600 analyzer? Only a small change is required from 0-50 ppm CO to 0-60 ppm.
A: Instructions are as follows:Model 7600 Range Change: Enter Maintenance Mode Select Item 7 Other Parameter Select Range Setting. Select the Channel / Component to be modified. This should bring up a screen that shows Min RangeRange 1Range 2Max RangeRange Num: Use the menu to adjust the values as needed.