The LXT-800 utilizes nephelometric sensing technology combined with the functionality of the proven LXT-220 style controller to provide reliable, continuous turbidity measurement. The optical sensor consists of an LED light source trans- mitting near-infrared light into the sample. Suspended particles present in the water scatters the light which is detected 90° from transmission, in accordance with ISO 7027 / EN 27027. Teledyne's design consists of a reference detector and two measuring detectors, which utilize different path lengths to provide a wider dynamic measuring range and high sensitivity on the low end. The LXT-800 also ensures balanced sensitivity regardless of particle size. The LXT-800 sensor utilizes a self-cleaning design of the sapphire window under moderate flow conditions. Sensor unit is available in both an immersion and flow-through configuration. The LXT-220 controller is weatherproof NEMA-4X ½ DIN form factor with universal mounting bracket available for panel, wall, handrail or pipe mounting. The membrane keypad and backlit LCD display provides an easy-to-use interface with menu-driven functionality. 4-20mA output is provided as standard and configurable alarm contacts can be provided optionally. Information is digitally communicated between sensor and control units which ensures EMI/RFI robustness and enables separation distances up to 200 meters. Applications
|
|||||||||
|
Document Control & Training (password) e-mail us: ask_tai@teledyne.com Copyright © 2010 Teledyne Technologies Incorporated. All rights reserved. Teledyne Analytical Instruments 16830 Chestnut St City of Industry, CA 91748 USA |