Frequently Asked Questions
For additional questions, please contact Technical Support.
Q: Linearization Curves for Thermal Conductivity Analyzers
A:
RANGE : 0-50% H2 BALANCE ARGONRANGE : 0-50% H2 BALANCE NITROGENRANGE : 0-50% HE BALANCE ARGONRANGE : 0-50% HE BALANCE NITROGEN
RANGE : 0-50% H2 BALANCE ARGON
DPT
Input
Output
0
10.62
5.00
1
19.47
10.00
2
26.21
15.00
3
31.77
20.00
4
36.29
25.00
5
39.99
30.00
6
43.08
35.00
7
45.76
40.00
8
48.06
45.00
Back to Top
RANGE : 0-50% H2 BALANCE NITROGEN
9.85
17.21
23.36
28.80
33.23
37.50
40.95
44.00
46.96
Back to Top
RANGE : 0-50% HE BALANCE ARGON
7.61
14.92
21.37
27.13
32.10
36.45
40.42
47.16
RANGE : 0-50% HE BALANCE NITROGEN
6.80
13.40
19.45
30.15
34.75
39.05
42.95
46.60
45.00
Back to Top
Q: How do you replace the T/C cell in the model 2020?
A: Please refer to the following documents:2020_TC_cel_replacement.pdfD67135-6_Model_2020_Final.pdf
Back to Top
Q: We are doing voltage drop calculations for some remote 2020H analyzers. The specifications state the voltage as 120 VAC but no amp load or wattage. Can you provide this information?
A: The maximum power of the unit is 200W.
Back to Top