2120XL FAQ​


​What is the effect of moisture on the nitrogen reading?
​Can the analyzer be powered without flowing Argon sample gas?
​Can oxygen or other impurities interfere with the nitrogen measurement?
​​What is the maximum sample inlet pressure?

For additional questions, please contact Technical Support​​.

Q: ​What is the effect of moisture on the nitrogen reading?

A: Without Teledyne's moisture trap, interference due to moisture is expected to reduce the nitrogen reading by approximately 4% of the reading for every ppm of moisture in the stream. With Teledyne's moisture trap installed, there will be no interference. ​

​​Back to Top


Q: ​Can the analyzer be powered without flowing Argon sample gas?

A: It is not recommended to keep the unit powered without flowing sample gas. It is possible detector damage could occur. ​

​​Back to Top​​​​


Q: Can oxygen or other impurities interfere with the nitrogen measurement?

A: Although the plasma emission detector for the 2120XL is nitrogen-selective, other compounds including nitrogen may introduce some small positive or negative interference. The interference is non-linear over the analyzer's ranges and may be depending on the impurity and concentration. 


Here is an example of interferences contributed by the impurities of a gas containing 10 ppm ​nitrogen, 10 ppm hydrogen, 10 ppm oxygen, 10 ppm methane and 10 ppm carbon monoxide in an argon background: 

Impurity

Contribution to Reading

% of Nitrogen Response​

Nitrogen

10 ppm

---

Oxygen

0.35 ppm

3.5%

Hydrogen

-0.2 ppm

-2.0%

Methane

-0.175 ppm

-1.75%

Carbon Monoxide

0.175 ppm

1.75%

​​Back to Top​​​​


Q: ​What is the maximum sample inlet pressure?​

A: 30 psig, although it is recommended not to exceed 20 psig for the analyzer to properly control flow rate. ​

​​​Back to Top​​​​​

​​