Gas Analyzers
The Model 3010C Series BASEEFA/CENELEC approved, intrinsically safe trace oxygen analyzers are versatile, microprocessor based instruments used for detecting oxygen from 0-10 ppm to 100% in a variety of gases.
Maintenance-Free
The series features a Micro-fuel Cell to measure oxygen in the 0-10 ppm to 100% range. The sensors set industry standards for accuracy, sensitivity and ease of use, and come with a one year limited warranty against defects in material and workmanship. Also, because every sensor is manufactured using stringent quality control procedures, the user is assured of outstanding reliability and optimum performance. The sensor is a sealed electrochemical device with no electrolyte to change or electrodes to lean, making it maintenance free. The sensor is specific to oxygen and can be used in a variety of background gases including hydrocarbons.
Flexibility
These units offer three dynamic, user-configurable ranges and can be calibrated at any range. Two fully programmable concentration alarms and relays provide the versatility to satisfy nearly any requirement.
Convenience
Two standard 0-1 VDC outputs and two 4-20mADC isolated outputs provide both concentration and range identification. A bi-directional RS-232 serial interface is incorporated to relay information to a host computer for remote monitoring of zero and span calibration.
Split Architecture
These units are designed as “split architecture” instruments, meaning that a general purpose control unit for non-hazardous areas controls a specially designed analysis unit or “remote probe” that is CENELEC approved intrinsically safe for use in area classification EEXiallCT6, Zone 0. Simple menu choices, membrane command switches and dual displays make setup and operation clear and quick. The 3010TC series is constructed of high quality materials to insure long term, trouble free materials to insure long term, trouble free service.
Options
K 19" rack mount with either one or two units installed, mounted side by side (flush mount versions only)
S Stainless steel cell holder
Alternate Sensors
Class L2C-L: Backgrounds with CO2 90% of FS in < 60 seconds; up to 1 year life
Maintenance-free Micro-Fuel Cell
Three user-selectable ranges plus cal range
Signal output: 0-1 VDC & 4-20 mADC
Programmable auto ranging
Range ID contacts (qty 4), Form A normally open contacts, 3A resistive
Two fully-adjustable concentration alarm points with programmable relay function, Form C contacts, 3A resistive
Programmable auto cal/zero with Form A normally open contact relay signals
Remotely indicated cal/zero via customer supplied 24 VDC signal and 12 VDC control valves
Self diagnostics with Form C failure alarm contacts
Full duplex RS-232 communication link
Five digit oxygen concentration LED display
Back-lit 2 x 20 line alphanumeric vacuum fluorescent display for setup and diagnostics
Universal power supply: 85-230 VAC, 50/60 Hz
Remote probe, nylon with stainless steel fittings
Oxygen sensor:
3010TC Micro-Fuel Cell, Class L2C, up to 2 years life
3010PC Micro-Fuel Cell, Class C-3, up to 1.5 years life
Ranges
3 customer selectable ranges3010TC Min: 0-10 ppm Max: 0-25%3010PC Min: 0-1% Max: 0-100%Calibration Range: 0-25% (unit can be calibrated on any range or gas)
Accuracy
± 2% of FS at a constant temperature± 5% of FS over operating temperature range (once temperature equilibrium has been reached; except 0-10 ppm scale which is ± 1 ppm)
Sensitivity
0.5% of FS
Response Time
3010TC: < 60 seconds3010PC: < 30 seconds
Temperature
32° to 122°F (0° to 50°C)
Output
Analytical: 0-1 VDC and 4-20 mADC isolated, 1000 ohm impedanceRange ID: 0-1 VDC and 4-20 mADC isolated
Power
83 to 230 VAC, 50/60 Hz
Enclosure
3010TAC: General purpose, flush panel mount3010TBC: I.P. 65/66 rated wall mount
Dimensions
3010TAC & PAC: Case: 8.7"W x 6.96"H x 12.2"D (20.1 x 17.68 x 30.99 cm) Bezel: 10.79"W x 7.46"H (27.4 x 18.95 cm)3010TBC & PBC: 25.63"H x 15.25"W x 11"D (65.1 x 38.74 x 27.94 cm)
Air separation and liquefaction
Pure gaseous hydrocarbon stream monitoring
Semiconductor manufacturing
Protective atmosphere blanketing of primary liquid feedstocks and flammable liquids
Process monitoring of gaseous monomers, vinyl chloride, propylene, butadiene, isoprene or ethylene
Gas purity certification