Model 3190

Trace Oxygen Analyzer​​​​


Description

The 3190 analyzer proves that good things come in small packages. This microprocessor based unit offers high accuracy, easy use and all the standard features demanded by the end-user in a compact, budget priced design. Membrane command switches and a large, four digit LED display make setup and operation clear and quick.


Range Options

Depending on the application, this unit can be configured to use a Micro-fuel Cell sensor or zirconium oxide sensor, assuring the highest accuracy. With a long expected sensor life, the 3190 is as inexpensive to maintain as it is to purchase. It has a sensor failure alarm which takes away the worry and work out of the end-user’s hands. Two ranges are field configurable at 0-10 ppm and 0-1000 ppm with a 0-25% calibration range. Optional ranges can be set between 10 ppm and 10,000 ppm. Two programmable failsafe concentration alarms (one high and one low set-point) provide the versatility to satisfy nearly any requirement.


Outputs

A standard 0-10 VDC output provides range identification, while the oxygen concentration is output through 4-10 mADC negative ground and 0-10 VDC negative ground signals. An optional unidirectional RS-232 serial interface is incorporated to relay information to a host computer for remote monitoring of all functions. Configuration information and analysis results  are as close as your personal computer.​


Oxygen Sensor

An SP76 compliant oxygen sensor for trace and percent ranges is offered with the 3190. This can substantially reduce size, as compared to the alternative that requires discrete tubing cut and connected through fittings of various components like sensors, pressure regulators, valves, etc.​ Additionally, it is a building block, plug-and-play approach that makes even the most inexperienced assembler or designer an immediate expert.


Additional Advantages

  • Auto-ranging capabilities

  • One high and one low alarm set-point with corresponding relay contacts

  • Sensor failure alarm


Optional Configurations

  • Power requirement: 10-36 VDC

  • Zirconium oxide sensor



Features

  • Principle: Oxygen specific* Micro-fuel Cell B2C

  • Ranges: Two ranges, with auto-ranging capability; field configurable between 0-10 ppm and 0-10,000 ppm O2

  • Calibration range: 0-25% O2

  • Signal outputs:

  • For oxygen concentration: 4-20 mADC negative ground and 0-10 VDC signals

  • For range identification: 0-10 VDC signal

  • ​Alarm outputs:

  • One high alarm and One low alarm set points, with corresponding relay contacts.

  • One sensor failure alarm

  • Display: 4 digit LED

  • Power requirements: Universal AC power supply 85-265 VAC, 47-63 Hz; 10 watts​


Specifications

​Ranges

​0-10 ppm and 0-1,000 ppm and 0-15% calibration range (nominal)
(user selectable percent ranges)
User selectable ranges can be set between 10 and 10,000 ppm

​Display

​LED

​Accuracy

​± 1% of FS at constant temperature

​Response Time

​90% in less than 45 seconds at 77°F (25°C) for fuel cell, faster with zirconium oxide sensor

​Temperature

​32° to 122°F (0° to 50°C)

​Signal Output

​4-20 mA and 0-10 VDC negative ground, non-isolated

​Range Identification

0-10 VDC

​Alarm

​One high alarm relay, adjustable
One low alarm relay, adjustable
One sensor calibration fail relay (all alarms are failsafe)

​Power

​10-36 VDC or 100V - 240VAC; 47-440 Hz, user specified

​Dimensions

​2.81"H x 6"W x 2.87"D
(7.14 x 15.24 x 7.29 cm)

​Oxygen Sensor

​Class B-2 (for ranges lower than 0-200)
Optional: Class A-2C for CO2 backgrounds
                  Zirconium oxide


Applications

  • Wave solder re-flow ovens

  • Glove box manufacturers

  • Nitrogen membrane skids

  • PSA skids

  • Helium recovery systems​

  • Diving applications

  • Clean air systems for medical applications

Downloads

Title​​​

​File Size​

Model 3190 Brochure

1.64MB

Model 3190 Manual

1.19MB

Model 31​90Z Manual

​453KB

Model 3190 Declaration of Conformity​

​​330K​B


​​