Learn about the ​2000RS

​​Model 3000RS Series​​​​

Oxygen Analyzers​​​

 

​​

Description

​The Model 3000RS Series is the next generation of online percent and trace oxygen analyzers from Teledyne Analytical Instruments. These analyzers are designed with the improved precision, sensitivity and accuracy of their predecessor, modern electronics and updated graphic user interface, all packaged into a compact and elegant chassis. 

Model 3000RS-P

The 3000RS-P offers 0-1%, up to 100%, oxygen analysis range using Teledyne's unique micro-fuel cell technology.  The micro-fuel cell absorbs and reacts with oxygen (O2) to generate a linear electrical signal, proportional to the concentration of O2 in the measured sample. The O2 sensor is robust and suitable for operation in a diverse gas background, and is unaffected by changes in sample flow.


Model 3000RS-T

The 3000RS-T offers 0-10ppm, up to 10,000ppm, oxygen analysis range using Teledyne's unique micro-fuel cell technology.  The micro-fuel cell absorbs and reacts with oxygen (O2) to generate a linear electrical signal, proportional to the concentration of O2 in the measured sample. The O2 sensor is robust and suitable for operation in a diverse gas background, and is unaffected by changes in sample flow. Teledyne's patented InstaTrace sensor shields the sensing membrane from high ambient oxygen exposure, resulting in a much faster installation and start up recovery time of new installed O2 sensors. 


Model 3000RS-TXL

The 3000RS-T offers 0-1ppm, up to 10,000ppm, oxygen analysis range using Teledyne's unique micro-fuel cell technology.  The micro-fuel cell absorbs and reacts with oxygen (O2) to generate a linear electrical signal, proportional to the concentration of O2 in the measured sample. The O2 sensor is robust and suitable for operation in a diverse gas background, and unaffected by changes in sample flow. Teledyne's patented InstaTrace-XL sensor shields the sensing membrane from high ambient oxygen exposure, resulting in a much faster installation and start up recovery time of new installed O2 sensors. 



Features

  • Mount Style: Bench Top or Flush Panel (specify at time of order)

  • Signal output: 0-1 VDC and 4-20 mADC (isolated)

  • 0-1 VDC and 4-20 mA (isolated) range identification

  • Programmable auto-ranging

  • Range ID contacts (qty 4) Form A normally open contacts

  • Two fully adjustable oxygen concentration alarm set points with programmable relay function Form C contacts

  • Programmable auto-calibration capability

  • Remotely operated calibration capability

  • Self-diagnostics with Form C failure alarm contacts

  • Profibus-DP and RS232 bi-directional communications

  • 3.5 inch color touchscreen LCD displays reading and key parameters

  • Integral MEMS-based flow sensor

  • RFI protected

  • Power: 24 VDC

  • CE Mark: Conforms to EN61326:2013 and EN61010-1:2010

  • Power adapter: 60W power supply (AC power cord ordered separately)


3000RS-P Only

  • Fast response B-1 micro-fuel cell (MFC) sensor​

  • Three user-selectable ranges from 0-1% to 0-100% O2


3000RS-T Only

  • ​Fast recovery InstaTrace micro-fuel cell (MFC) sensor (for CO2 applications, use InstaTrace-CO2)

  • Three user selectable ranges from 0-10 ppm up to 0-10,000 ppm


3000RS-TXL Only

  • ​Fast recovery InstaTrace-XL micro-fuel cell (MFC) sensor with InstaTrace cell holder

  • Three user selectable ranges from 0-1 ppm up to 0-10,000 ppm​

Specifications

​​
​​
​3000RS-P
​​3000RS-T
​Sensor
​Micro-Fuel Cell
​Micro-Fuel Cell
​Ranges
​​0-1% to 0-100%
​0-10 ppm to 0-10,000 ppm
(0-1 ppm with -XL option)​
​Approvals
​CE marked​
​Repeatability
​1% FS
​1% FS
​Linearity
​1% FS
​1% FS
​Response Time
​< 10 seconds
​< 10 seconds for 0-1,000 ppm range
< 60 seconds for 0-10 ppm range
< 100 seconds for 0-1 ppm range(-XL only)
​Temperature
​32° to 122°F (0° to 50°C)
​A-Class Sensor: 32° to 104°F (0° to 40°C)
B-Class Sensor: 32° to 104°F (0° to 50°C)
​Inlet Flow
​1-2 SCFH (0.5-1.0 SLPM)
​1-2 SCFH (0.5-1.0 SLPM)
Display
​3.5" touchscreen color LCD​
​​I/O
​​4-20 mADC (isolated) or 0-1 VDC analog outputs
1 x system fault alarm, Form C relay contact
2 x concentration alarms, Form C relay contact
4-20 mADC (isolated) or 0-1 VDC range ID
4 x Form A (normally open) Range ID contacts
Profibus-DP and RS232 communications
​Gas Connections
​1/8" compression fittings for all connections
​Voltage
​24 VDC (optional 100-240 VAC, 50/60 Hz with AC adapter)
​Power
​10 W Nominal
​10 W Nominal
​Dimensions
​4.3"W x 5.1"H x 17.8"L (10.9 x 12.9 x 45.2 cm)

Applications

  • Air separation and liquefaction

  • Pure, gaseous hydrocarbon stream monitoring

  • Emissions monitoring

  • Protective atmosphere blanketing of primary liquid feedstocks and flammable liquids

  • Process monitoring of gaseous monomers - vinyl chloride, propylene, butadiene​, isoprene or ethylene

  • Gas purity certification

  • Semi conductor manufacturing