​​235 FAQ​​

​Linearization Curves for Thermal Conductivity Analyzers
​Our old 235 unit is drifting up scale all the time. Power supplies are good, block temperatures are good and stable, zero and span gases have not been changed, but there is a slight imbalance on the bridge resistance. Is the cell block bad?
​What effect does sample and reference flow rate have on the reading when the unit is configured to measure 0-10 N2 in Argon?

For additional questions, please contact Technical Support​


Q: ​​​Linearization Curves for Thermal Conductivity Analyzers

A:  

RANGE : 0-50% H2 BALANCE ARGON
RANGE : 0-50% H2 BALANCE NITROGEN
RANGE : 0-50% HE BALANCE ARGON
RANGE : 0-50% HE BALANCE NITROGEN

RANGE : 0-50% H2 BALANCE ARGON

DPT

Input

Output

0

10.62

5.00

1

19.47

10.00

2

26.21

15.00

3

31.77

20.00

4

36.29

25.00

5

39.99

30.00

6

43.08

35.00

7

45.76

40.00

8

48.06

45.00

​​Back to Top​​​​​


RANGE : 0-50% H2 BALANCE NITROGEN

DPT

Input

Output

0

9.85

5.00

1

17.21

10.00

2

23.36

15.00

3

28.80

20.00

4

33.23

25.00

5

37.50

30.00

6

40.95

35.00

7

44.00

40.00

8

46.96

45.00

​​Back to Top​​​​​


RANGE : 0-50% HE BALANCE ARGON

DPT

Input

Output

0

7.61

5.00

1

14.92

10.00

2

21.37

15.00

3

27.13

20.00

4

32.10

25.00

5

36.45

30.00

6

40.42

35.00

7

44.00

40.00

8

47.16

45.00

​​Back to Top​​​​​


RANGE : 0-50% HE BALANCE NITROGEN

DPT

Input

Output

0

6.80

5.00

1

13.40

10.00

2

19.45

15.00

3

25.00

20.00

4

30.15

25.00

5

34.75

30.00

6

39.05

35.00

7

42.95

40.00

8

46.60

45.00​

​​Back to Top​​​​​


Q: ​​​Our old 235 unit is drifting up scale all the time. Power supplies are good, block temperatures are good and stable, zero and span gases have not been changed, but there is a slight imbalance on the bridge resistance. Is the cell block bad?

A:  Old 235 units normally go bad when the thermistor that goes into the cell block degrades from years of being at a high temperature. The insulation dries​​ out and the whole thermistor assembly tends to fall apart.

Also, you can check the stability of the voltage drive directly at the bridge circuit. Make sure this is steady and the correct number.

Also, be certain it isn't a process change, or a leak in the tubing somewhere.  What about flow changes -- any possibility of large changes in flow rate or sample pressure?

It is rare for the cells to go bad, except for maybe sealed references.​

​​Back to Top​​​​​


Q: ​​​What effect does sample and reference flow rate have on the reading when the unit is configured to measure 0-10 N2 in Argon?

A:  The effect of sample and reference flow rate change in the range mentioned below is minimal. This data applies:


Sample Flow Rate

Reference Gas Flow Rate

Analyzer Reading Percent N2

1000 cc/m

0.5 scfh

0.03

1000 cc/m

0.025

0.03

1000 cc/m

1

0.03

500 cc/m

1

-0.01

250 cc/m

1

-0.02​

​​Back to Top

​​